2858
:SENSe:AVERage:COUNt
Type
Command/Query
Syntax
:SENSe:AVERage:COUNt <numeric>
:SENSe:AVERage:COUNt?
Description
This command sets averaging factor value.
Parameter
Description Number of Averaging factor.
Data Type Long integer type (Long)
Range 1 to 999
Preset Value
16
Unit -
Resolution -
Query Response
{numeric}<newline><^END>
Examples
10 OUTPUT 717;":SENS:AVER:COUN 16"
20 OUTPUT 717;":SENS:AVER:STAT ON"
Related Commands
:SENSe:AVERage:STATe
:SENSe:AVERage:CLEar
Equivalent GUI
Setup > Averaging > Type in the average factor in to entry box beside
Averaging