E5071C
2872
:SENSe:CORRection:RIMPedance
Type
Command/Query
Syntax
:SENSe:CORRection:RIMPedance <numeric>
:SENSe:CORRection:RIMPedance?
Description
This command sets reference impedance value.
Parameter
Description Reference impedance value
Data Type Double precision floating point type (Double)
Range 1m to 10M
Preset Value
50
Unit -
Resolution -
Query Response
{numeric}<newline><^END>
Examples
10 OUTPUT 717;":SENS:CORR:RIMP 0.001"
20 OUTPUT 717;":SENS:CORR:RIMP?"
30 ENTER 717;A
Related Commands
:SENSe:CORRection:DCONstant
Equivalent GUI
Setup > Ref. Z (under More Functions)