Programming
2877
TRIGGER
:TRIGger:AVERage
Type
Command/Query
Syntax
:TRIGger:AVERage {ON|OFF|1|0}
:TRIGger:AVERage?
Description
This command sets averaging trigger on/off. When averaging trigger is on,
one trigger makes one averaging measurement. For example, if the
averaging factor is set at 16, one trigger makes a measurement for 16
times.
Parameter
Description Status of A
veraging trigger
Data Type Boolean type (Boolean)
Range ON|OFF|1|0
Preset Value
OFF
Unit -
Resolution -
Query Response
{1|0}<newline><^END>
Examples
10 OUTPUT 717;":TRIG:MODE HOLD"
20 OUTPUT 717;":SENS:AVER:COUN 16"
30 OUTPUT 717;":SENS:AVER:STAT ON"
40 OUTPUT 717;":TRIG:AVER ON"
50 OUTPUT 717;":TRIG:SING"
Related Commands
:SENSe:AVERage:COUNt
:SENSe:AVERage:STATe
:TRIGger:SINGle
Equivalent GUI

Table of Contents