Programming
2879
:TRIGger:MODE
Type
Command/Query
Syntax
:TRIGger:MODE {HOLD|SINGle|RUN}
:TRIGger:MODE?
Description
This command sets trigger mode.
Parameter
Description Trigger mode
Data Type Character string type (String)
Range
HOLD: Trigger is on hold until the conditions are meet,
the trigger event starts.
SINGle: Trigger event is run once.
RUN: Trigger event is running continuous.
Preset
Value
RUN
Unit -
Resolution -
Query Response
{HOLD|SINGle|RUN}<newline><^END>
Examples
10 OUTPUT 717;":TRIG:MODE HOLD"
20 OUTPUT 717;":TRIG:MODE?"
30 ENTER 717;A$
Related Commands
:TRIGger:SINGle
Equivalent GUI
Stop Single or Run