E5071C
348
Unsupported
Version
This message is displayed when the state file generated
from a newer version of Firmware is recalled.
Invalid No. of
Ports
This message is displayed when the state file generated
from a Firmware having different Port Options is recalled.
Source
Attenuator
Option
Mismatch
This message is displayed when the state file generated
from a Firmware having different Source Attenuator
Option is recalled.
Power Sweep
Option
Mismatch
This message is displayed when the state file generated
from a Firmware having different Power Sweep Option is
recalled.
Invalid No. of
Points
This message is displayed when the Maximum No. of
Points are not enough.
External Test
Set Option
Mismatch
This message is displayed when the state file generated
from a Firmware having different External Test Set Option
is recalled.
Frequency
Option
Mismatch
This message is displayed when the state file generated
from a Firmware having different Frequency Option is
recalled.
ch x, tr y, z pts
required
This message is displayed when the Recall Function cannot
recall data with maximum channel or traces. The number
of channels/traces should be higher than the recall state
file to recover from the error. More information is available
in Setting upper limit of channels and traces.
• The warning message listed above:
1. May not appear in all failures
2. Appears on the screen only when recall is done from soft key
by Recall State or Recall by File Name.
3. Displays "Recall Failed" once when the recall is done through
GPIB or VBA.
4. Is not kept in Error queue. This means that when
SCPI.SYSTem.ERRor is executed, "Recall Failed" is returned.
5. These messages are supported for state files created in
Firmware revision A.09.10 or later.
• When a user file is used in Extending the Calibration Plane Using
Network De-embedding, Determining Characteristics After
Adding a Matching Circuit, or Determining the Characteristics
that Result from Adding a Matching Circuit to a Differential Port
and the setup status is saved, a recall error will occur if the user

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers: