Product Information
781
rounded exceeds the range of -65536 to +65536
or when a real-number-based command for which
the parameter can be rounded exceeds the range
of -9.9e37 to +9.9e37, for example.
-
104
Data type
error
The parser has recognized a data element that
must not exist. Block data has been sent instead
of numeric value data or character string data
that had been expected, for example.
E
No. Message Description
-
200
Execution
error
An error associated with execution has been
generated for which this instrument cannot
specify the error message. This code shows that
an error associated with execution defined in
11.5.1.1.5, IEEE488.2 has occurred. This error
occurs also when a calibration measurement is
aborted.
-
123
Exponent
too large
The absolute value of the exponent exceeds
32,000 (see 7.7.2.4.1, IEEE488.2).
-
178
Expression
data not
allowed
An expression-data element has been received at
a position where this instrument does not accept
one.
-
170
Expression
error
When the expression data is put to syntactic
analysis, an error not corresponding to one of
Error Numbers -171 through -179 occurs.
F
No. Message Description
100
Failed to
read file
This error occurs when load eye-mask file
(:MMEM:LOAD:EYE:MASK) and load user defined
bit pattern file (:MMEM:LOAD:EYE:BPAT) cannot
be read correctly.
101
Failed to
write file
This error also occurs when formatted data
(:MMEMy:STOR:FDAT), displayed image
(:MMEM:STOR:IMAG) and measurement data
(:MMEM:STOR:SNP) cannot be stored correctly.