Product Information
845
1. Turn on the E5071C.
2. Press Macro Setup key.
3. Click Load Project... in the softkey menu.
4. Select RestoreSysCorFile.vba from the D:\Agilent\Service folder and
then press the Open button.
5. Press Macro Run key
6. The RestoreSysCorFile dialog box appears.
7. Click OK.
8. Once the program has terminated, restart the E5071C to enable
the system correction data.