Programming
1051
phase
To read/write one of the formatted data arrays, use the following
command:
:CALC{1-160}:DATA:FDAT
Formatted memory arrays
A formatted memory array contains the formatted data (values to be
displayed) obtained by performing data math operations, measurement
parameter conversion, and smoothing on a particular corrected memory
array.
To read/write one of the formatted memory arrays, use the following
command:
:CALC{1-160}:DATA:FMEM
Stimulus data arrays
A stimulus data array contains the stimulus values for all measurement
points.
The instrument retains 160 stimulus data arrays at maximum, each of
which is associated with one of the 160 channels. Stimulus data arrays are
read-only. To retrieve one of the stimulus data arrays, use the following
command:
:SENS{1-160}:FREQ:DATA?