Programming
1101
a command) represented with High(0)/Low(1) of each bit of the port can
be output.
To output data, use the following command:
Port Name Command
Port A
:CONT:HAND:A
Port B
:CONT:HAND:B
Port C
:CONT:HAND:C
Port D
:CONT:HAND:D
Port E
:CONT:HAND:E
Port F
:CONT:HAND:F
The bit 6 of the data output by :CONT:HAND:B (the bit 14 of the
data output by :CONT:HAND:F) is ignored when outputting the
/INDEX signal is turned on.
The bit 7 of the data output by :CONT:HAND:B (the bit 15 of the
data output by :CONT:HAND:F command) is ignored when
outputting the /READY FOR TRIGGER signal is turned on.
Sample Program
See Handler Interface.