Programming
1181
1720 OUTPUT @Agte507x;"*OPC?"
1730 ENTER @Agte507x;Buff$
1740 PRINT "Set LOAD to Port "&VAL$(Port(I))&". Then push [Enter] key."
1750 INPUT "",Buff$
1760 OUTPUT @Agte507x;":SENS"&Ch$&":CORR:COLL:LOAD ";Port(I)
1770 OUTPUT @Agte507x;"*OPC?"
1780 ENTER @Agte507x;Buff$
1790 NEXT I
1800 !
1810 ! Transmission Measurement
1820 !
1830 FOR I=1 TO Num_of_ports-1
1840 FOR J=I+1 TO Num_of_ports
1850 PRINT "Set THRU between Port "&VAL$(Port(I))&" and Port "& VAL$(Port(J))&". Then push [Enter]
key."
1860 INPUT "",Buff$
1870 OUTPUT @Agte507x;":SENS"&Ch$&":CORR:COLL:THRU ";Port(I);"," ;Port(J)
1880 OUTPUT @Agte507x;"*OPC?"
1890 ENTER @Agte507x;Buff$
1900 OUTPUT @Agte507x;":SENS"&Ch$&":CORR:COLL:THRU ";Port(J);"," ;Port(I)
1910 OUTPUT @Agte507x;"*OPC?"
1920 ENTER @Agte507x;Buff$
1930 NEXT J
1940 NEXT I
1950 !
1960 ! Done
1970 !
1980 OUTPUT @Agte507x;":SENS"&Ch$&":CORR:COLL:SAVE"
1990 PRINT "Done"