E5071C
192
3. Select STD Type > Delay/Thru.
4. Set Offset Loss to 1.3Gohm/s, and Offset Delay to 0.
5. Click Return to return to the Define STD menu.
6. In the same way, repeat the procedure to enter the definition of
REFLECT to No. 2.
7. Select SHORT for STD Type.
Procedure to define Match
1. Click 3:No Name > Label.
2. Type in MATCH <2G.
3. Click STD Type > Load.
4. Set Max Frequency to 2GHz.
5. Click Return to return to the Define Std menu.
Procedure to define Line 1/2
1. Click 4:No Name - Label.
2. Type in LINE <7G.
3. Click STD Type - Thru.
4. Set Offset Delay to 54 ps.
5. Set Min Frequency to 2 GHz.
6. Set Max Frequency to 7 GHz.
7. Click Return to return to the Define Std menu.
8. Click 5:No Name - Label.
9. Type in LINE >7G.
10. Click STD Type > Thru.
11. Set Offset Delay to 13 ps.
12. Set Min Frequency to 7 GHz.
13. Click Return to return to the Define Std menu.
14. Click Return to return to the Modify Cal Kit menu.
Procedure to register Thru/Reflect/Match to sub class 1
1. Click Specify CLSs > Sub Class.
2. Select Sub Class1.
3. Click TRL Thru > Set All > THRU > Return.
4. Click TRL Reflect > REFLECT.
5. Click TRL Line/Match > Set All > MATCH <2G > Return.

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals