E5071C
2772
SCPI.TRIGger.SEQuence.IMMediate
Object type
Method (Write-only)
Syntax
SCPI.TRIGger.SEQuence.IMMediate
Description
This command generates a trigger immediately and executes a
measurement, regardless of the setting of the trigger mode.
This command is different from SCPI.TRIGger.SEQuence.SINGle as the
execution of the object finishes at the time of a trigger.
If you execute this object when the trigger system is not in the
trigger wait state (trigger event detection state), an error occurs
when executed and the object is ignored.
Examples
SCPI.TRIGger.SEQuence.SOURce = "bus"
SCPI.INITiate(1).CONTinuous = True
SCPI.TRIGger.SEQuence.IMMediate
Related objects
SCPI.TRIGger.SEQuence.IMMediate
Equivalent key
No equivalent key is available on the front panel.
Equivalent SCPI command
Syntax
:TRIGger[:SEQuence][:IMMediate]
Example of use
10 OUTPUT 717;":TRIG:IMM"
20 OUTPUT 717;"*OPC?"
30 ENTER 717;A