E5071C
2778
SCPI.TRIGger.SEQuence.SOURce
Object type
Property (Read-Write)
Syntax
SCPI.TRIGger.SEQuence.SOURce = Param
Param = SCPI.TRIGger.SEQuence.SOURce
Description
This command sets/gets the trigger source from the following 4 types:
1. Internal Trigger
Uses the internal trigger to generate continuous triggers automatically.
2. External Trigger
Generates a trigger when the trigger signal is inputted externally via
the Ext Trig connector or the handler interface.
3. Manual Trigger
Generates a trigger when the key operation of Trigger > Trigger is
executed from the front panel.
4. Bus Trigger
Generates a trigger when the SCPI.IEEE4882.TRG object is executed.
5. When you change the trigger source during sweep, the sweep is
aborted.
Variable
Parameter Param
Description Trigger source
Data type Character string type (String)
Range
Select from either of the following:
"INTernal": Specifies internal
trigger.
"EXTernal": Specifies external
trigger.
"MANual": Specifies manual
trigger.
"BUS": Specifies bus trigger.