2. Pressing the Execute (3 in menu) button executes a phase error
correction and overwrites the results on the original data file.
• You cannot run the Execute function when selecting the
measurement port if the data file of the calibration mixer with IF
filter (*.s2p) has not been set for the network removal function
of the fixture simulator.
• The phase error correction data reflects the phase difference of
the LO signals for the phase information of the calibration
mixer's data file, which is registered in any two ports.
3. Press the Close button (4 in menu) to exit macro.