E5071C
2804
:CALCulate:EYE:RESults:THReshold
Type
Command/Query
Syntax
:CALCulate:EYE:RESults:THReshold {T1_9|T2_8}
:CALCulate:EYE:RESults:THReshold?
Description
This command sets the rise time threshold level for the results of eye
measurement.
Parameter
Description Rise time threshold level
Data Type Character string type (String)
Range
T1_9: 10-90%
T2_8: 20-80%
Preset Value
T1_9
Unit -
Resolution -
Query Response
{T1_9|T2_8}<newline><^END>
Examples
10 OUTPUT 717;":CALC:EYE:RES:THR T1_9"
20 OUTPUT 717;":CALC:EYE:RES:THR?"
30 ENTER 717;A$
Related Commands
:CALCulate:EYE:RESults:DATA
Equivalent GUI
Eye Mask > Rise Time Def. (Under Result)