Programming
2809
:CALCulate:TRACe{Tr}:DTIMe:POSition
Type
Command/Query
Syntax
:CALCulate:TRACe{Tr}:DTIMe:POSition <numeric>
:CALCulate:TRACe{Tr}:DTIMe:POSition?
Description
This command sets delta time reference position.
Parameter
Description Delta time reference position.
Data Type Double precision floating point type (Double)
Range 0 to 100
Preset Value
50
Unit -
Resolution -
Query Response
{numeric}<newline><^END>
Examples
10 OUTPUT 717;":CALC:TRAC1:DTIM:POS 0"
20 OUTPUT 717;":CALC:TRAC1:DTIM:POS?"
30 ENTER 717;A
Related Commands
:CALCulate:TRACe{Tr}:DTIMe:DATA
:CALCulate:TRACe{Tr}:DTIMe:STATe
:CALCulate:TRACe{Tr}:DTIMe:TARGet
Equivalent GUI
Marker Search > Δ Time > Position(%) (In Δ Time Dialog box)