EasyManua.ls Logo

Agilent Technologies E5071C - Page 2581

Agilent Technologies E5071C
2685 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E5071C
2808
:CALCulate:TRACe{Tr}:DTIMe:DATA
Type
Query Only
Syntax
:CALCulate:TRACe{Tr}:DTIMe:DATA?
Description
This command gets delta time result value. You can get the result even if
:CALCulate:TRACe{Tr}:DTIMe:STATe is off.
Parameter
Description Value of delta time result.
Data Type Double precision floating point type (Double)
Range -
Preset Value
-
Unit -
Resolution -
Query Response
{numeric}<newline><^END>
Examples
10 OUTPUT 717;":CALC:TRAC1:DTIM:TARG 5"
20 OUTPUT 717;":CALC:TRAC1:DTIM:POS 50"
30 OUTPUT 717;":CALC:TRAC1:DTIM:DATA?"
40 ENTER 717;A
Related Commands
:CALCulate:TRACe{Tr}:DTIMe:POSition
:CALCulate:TRACe{Tr}:DTIMe:STATe
:CALCulate:TRACe{Tr}:DTIMe:TARGet
Equivalent GUI
No equivalent GUI is available.