E5071C
2834
:CALCulate:TRACe{Tr}:TIME:STEP:RTIMe:THReshold
Type
Command/Query
Syntax
:CALCulate:TRACe{Tr}:TIME:STEP:RTIMe:THReshold {T1_9|T2_8}
:CALCulate:TRACe{Tr}:TIME:STEP:RTIMe:THReshold?
Description
This command sets rise time threshold for the transform function.
Parameter
Description Transform function rise time threshold
Data Type Character string type (String)
Range
T1_9 : 10-90%
T2_8 : 20-80%
Preset Value
T1_9
Unit -
Resolution -
Query Response
{T1_9|T2_8}<newline><^END>
Examples
10 OUTPUT 717;":CALC:TRAC1:TIME:STEP:RTIM:THR T1_9"
20 OUTPUT 717;":CALC:TRAC1:TIME:STEP:RTIM:THR?"
30 ENTER 717;A$
Related Commands
:CALCulate:TRACe{Tr}:TIME:STEP:RTIMe:DATA
:CALCulate:TRACe{Tr}:TIME:STIMulus
Equivalent GUI
TDR/TDT > Parameters > 10-90% or 20-80% (In Stimulus)