Programming
2861
:SENSe:CORRection:COLLection:DLComp:LOAD
Type
Command Only
Syntax
:SENSe:CORRection:COLLection:DLComp:LOAD
Description
This command executes load measurement, as a part of Loss
Compensation sequence.
Parameter
Description Port number where the load measurement is executed.
Data Type Long integer type (Long)
Range 1 to 4
Preset Value
-
Unit -
Resolution -
Examples
10 OUTPUT 717;":SENS:CORR:COLL:DLC:LOAD 1 "
20 OUTPUT 717; ":SENS:CORR:COLL:DLC:SAVE"
Related Commands
:SENSe:CORRection:COLLection:DLComp:OPEN
:SENSe:CORRection:COLLection:DLComp:THRU
:SENSe:CORRection:COLLection:DLComp:SAVE
Equivalent GUI
Setup > Deskew & Loss Compensation (under Basic) > Measure > Port 1
and/or Port 2 and/or Port 3 and/or Port 4