E5071C
2864
:SENSe:CORRection:COLLection:DLComp:THRU
Type
Command Only
Syntax
:SENSe:CORRection:COLLection:DLComp:THRU
Description
This command executes a thru measurement, as a part of Loss
Compensation sequence.
Parameter
Description Port number where the thru measurement is executed.
Data Type Character string type (String)
Range
TH12: Thru between ports 1 and 2.
TH13: Thru between ports 1 and 3.
TH24: Thru between ports 2 and 4.
Preset Value
-
Unit -
Resolution -
Examples
10 OUTPUT 717;":SENS:CORR:COLL:DLC:THRU TH12"
20 OUTPUT 717; ":SENS:CORR:COLL:DLC:SAVE"
Related Commands
:SENSe:CORRection:COLLection:DLComp:OPEN
:SENSe:CORRection:COLLection:DLComp:LOAD
:SENSe:CORRection:COLLection:DLComp:SAVE
Equivalent GUI
Setup > Deskew & Loss Compensation (under Basic) > Measure
This applies to all other topologies except for Single-Ended
1-Port.