EasyManua.ls Logo

Agilent Technologies E5071C - Page 2637

Agilent Technologies E5071C
2685 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E5071C
2864
:SENSe:CORRection:COLLection:DLComp:THRU
Type
Command Only
Syntax
:SENSe:CORRection:COLLection:DLComp:THRU
Description
This command executes a thru measurement, as a part of Loss
Compensation sequence.
Parameter
Description Port number where the thru measurement is executed.
Data Type Character string type (String)
Range
TH12: Thru between ports 1 and 2.
TH13: Thru between ports 1 and 3.
TH24: Thru between ports 2 and 4.
Preset Value
-
Unit -
Resolution -
Examples
10 OUTPUT 717;":SENS:CORR:COLL:DLC:THRU TH12"
20 OUTPUT 717; ":SENS:CORR:COLL:DLC:SAVE"
Related Commands
:SENSe:CORRection:COLLection:DLComp:OPEN
:SENSe:CORRection:COLLection:DLComp:LOAD
:SENSe:CORRection:COLLection:DLComp:SAVE
Equivalent GUI
Setup > Deskew & Loss Compensation (under Basic) > Measure
This applies to all other topologies except for Single-Ended
1-Port.

Table of Contents

Other manuals for Agilent Technologies E5071C

Related product manuals