Programming
2869
:SENSe:CORRection:EXTension:AUTO:PORT{pt}
Type
Command/Query
Syntax
:SENSe:CORRection:EXTension:AUTO:PORT{[1]-4} {ON|OFF|1|0}
:SENSe:CORRection:EXTension:AUTO:PORT{[1]-4}?
Description
This command sets ON/OFF or gets the status of auto port extension.
Parameter
Description Status of auto port extension
Data Type Boolean type (Boolean)
Range ON|OFF|1|0
Preset Value
ON
Unit -
Resolution -
Query Response
{1|0}<newline><^END>
Examples
10 OUTPUT 717;":SENS:CORR:EXT:AUTO:PORT1 ON"
20 OUTPUT 717;":SENS:CORR:EXT:AUTO:PORT1?"
30 ENTER 717;A
Related Commands
:SENSe:CORRection:EXTension:AUTO:IMMediate
:SENSe:CORRection:EXTension:AUTO:STANdard
Equivalent GUI
Setup > Deskew (under Basic) > Option > Port1 and/or Port2 and/or Port3
and/or Port4 OR
Setup > Deskew & Loss Compensation (under Basic) > Option > Port1
and/or Port2 and/or Port3 and/or Port4