E5071C
462
A Touchstone data file is
defined for a single normalized
impedance value (real). Thus, when
preparing a Touchstone data file for the
matching circuit feature, be sure to
specify the same port reference
impedance value (real) to both 2-port
network in the file.
Once a user file is specified, the selection of Select
Circuit automatically changes to User.
6. For the circuit models, see Circuit models for defining matching
circuits.
7. Specify the values of the elements in the selected circuit model.
Softkey Function
C
Specifies capacitance [F]
G
Specifies conductance [S]
L
Specifies inductance [H]
R
Specifies resistance [ohm]
1. When either of the two components of Series-C, "C" and
"G", is set to 0, the component is defined as "Open";
when both components are set to 0, they are defined as
"Short."
When either of the two components of Shunt-L, "L" and
"R", is set to 0, the component is defined as "Short";
when both components are set to 0, they are defined as
"Open."
8. Repeat the procedure above to set up the matching circuit for each
port used.
9. Click Port Matching to turn the matching circuit function ON.
10. Click Return.
11. If Fixture Simulator is OFF, press the key again to turn it ON.

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals