Product Information
767
instrument is faulty. Contact an Agilent
Technology sales office or the company
from which you bought the instrument.
If this error occurs, the stimulus signal
output is automatically turned to off.
• For more information, see
Overload Detection and Power
Trip Function.
222
Port 2 receiver
overload
The input to Test Port 2 exceeds the
maximum input level. The measurement
value is not correct. When a DUT is an
amplifier or the like, this error may occur,
damaging the receiver in the worst case.
Should this error occur with a passive
part used as the DUT or with nothing
connected to the test port, this
instrument is faulty. Contact an Agilent
Technology sales office or the company
from which you bought the instrument.
If this error occurs, the stimulus signal
output is automatically turned to off.
• For more information, see
Overload Detection and Power
Trip Function.
223
Port 3 receiver
overload
The input to Test Port 3 exceeds the
maximum input level. The measurement
value is not correct. When a DUT is an
amplifier or the like, this error may occur,
damaging the receiver in the worst case.
Should this error occur with a passive
part used as the DUT or with nothing
connected to the test port, this
instrument is faulty. Contact an Agilent
Technology sales office or the company
from which you bought the instrument.
If this error occurs, the stimulus signal
output is automatically turned to off.
• For more information, see
Overload Detection and Power
Trip Function.