E5071C
768
224
Port 4 receiver
overload
The input to Test Port 4 exceeds the
maximum input level. The measurement
value is not correct. When a DUT is an
amplifier or the like, this error may occur,
damaging the receiver in the worst case.
Should this error occur with a passive
part used as the DUT or with nothing
connected to the test port, this
instrument is faulty. Contact an Agilent
Technology sales office or the company
from which you bought the instrument.
If this error occurs, the stimulus signal
output is automatically turned to off.
• For more information, see
Overload Detection and Power
Trip Function.
73
Power meter
not settled
The measurement value of the power
meter is unstable, or it goes outside the
tolerance.
This error occurs when the power meter's
measurement value does not become
stable within about 10 seconds during
measurement for power calibration data
or that for scalar-mixer calibration data.
This error also occurs when the power
calibration value exceeds the set
tolerance.
In addition, it occurs when the power
meter fails to respond for any reason,
such as power-down for example, during
measurement under the same conditions.
Ensure a firm connection of the power
sensor.
241
Power on test
failed
This error occurs when the power-on test
fails, indicating a failure of this
instrument. Contact an Agilent
Technology sales office or the company
from which you bought the instrument.
61
Power The out power level exceeds available