E5071C
770
-400 Query error
A comprehensive query error has
occurred showing that this instrument
cannot detect a more detailed error.
This code simply shows that a query
error defined in 11.5.1.1.7 and 6.3,
IEEE488.2 has occurred.
-410
Query
INTERRUPTED
The state that generates a
"INTERRUPTED" Query error (see
6.3.2.3, IEEE488.1). This error occurs
when data bytes (DAB) or GET are
received before the transmission of the
response after a query has not been
completed, for example.
-420
Query
UNTERMINATED
The state that generates an
"UNTERMINATED" Query error (see
6.3.2, IEEE488.2). This error occurs
when this instrument is designated as
the talker and an incomplete program
message is received, for example.
-440
Query
UNTERMINATED
after indefinite
response
After a query asking for an indefinite
response has been run, another query
is received in the same program
message (See 6.5.7.5.7, IEEE488.2).
-350 Queue overflow
The queue contains a specific code in
place of the code which caused this
error. The code indicates that the error
occurred because of no space available
in the queue, but the error is not
recorded.
R
No. Message Description
105 Recall failed
This error occurs when reading an
instrument status file (State01.sta,
etc.) (MMEM:LOAD:STAT command)
fails.
S
No. Message Description