Product Information
773
control is enabled.
-102 Syntax error
A command or data type that is not
recognized exists.
-310 System error
One of the errors designated as
"system errors" in this instrument has
occurred.
T
No. Message Description
40
Target value not
found
This error occurs when the target is
not found during the marker search
analysis after specifying the target
and executing the
CALC:MARK:FUNC:EXEC and
CALC:FUNC:EXEC commands. This
error occurs also when the bandwidth
is not found after executing the
bandwidth marker command,
CALC:MARK:BWID:DATA?
242
TDR option
requires
advanced
system
calibration
TDR requires the system information
PCC2.00 and above. This instrument
should be sent to service centre for
adjustment (with the firmware Rev
9.50 and above) to suit E5071C-TDR
option.
-124 Too many digits
The number of digits of the argument
of the decimal numeric-value-data
element exceeds 255 with the
preceding 0 removed (see 7.7.2.4.1,
IEEE488.2).
-223 Too much data
The block-, expression-, or character-
string-type program data that has
been received conforms with the
standard. But it exceeds the amount
that can be processed under the
condition of the memory or
conditions specific to memory-related
devices. In this instrument, this error
occurs when the number of

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals