Product Information
785
":CALC:DEV SEND1, SEND2" is sent
instead of a correct program message
with a command ":CALC:DEV SEND1"
which requires a parameter, the
instrument receives the message as the
number of parameters is invalid. See the
command reference to confirm the
required number of parameters.
-112
Program
mnemonic too
long
The length of the header exceeds 12
characters (see 7.6.1.4.1, IEEE488.2).
Q
No. Message Description
-430
Query
DEADLOCKED
The state that generates a
"DEADLOCKED" Query error (see
6.3.1.7, IEEE488.2). This error occurs
when both input and output buffers
have become full, preventing the
instrument from continuing processing,
for example.
-400 Query error
A comprehensive query error has
occurred showing that this instrument
cannot detect a more detailed error.
This code simply shows that a query
error defined in 11.5.1.1.7 and 6.3,
IEEE488.2 has occurred.
-410
Query
INTERRUPTED
The state that generates a
"INTERRUPTED" Query error (see
6.3.2.3, IEEE488.1). This error occurs
when data bytes (DAB) or GET are
received before the transmission of the
response after a query has not been
completed, for example.
-420
Query
UNTERMINATED
The state that generates an
"UNTERMINATED" Query error (see
6.3.2, IEEE488.2). This error occurs
when this instrument is designated as
the talker and an incomplete program