E5071C
60
7. Specify the IF bandwidth of the receiver as necessary. In this
measurement example, the IF bandwidth is set to 10 kHz because of
the need to lower the noise floor.
Avg > IF Bandwidth > 1 > 0 > k/m
STEP 2. Calibration
To turn the error correction ON, set the calibration type to the full 2-port
calibration and measure the calibration data.
For details about calibration, see Calibration.
1. Select the calibration kit suitable for the measurement cable. In this
measurement example, Calibration Kit 85032F is selected.
Cal > Cal Kit > 85032F
2. Set the calibration type to the full 2-port calibration using the test
port 1 and 2.
Cal > Calibrate > 2-Port Cal > Select Ports - 1-2
3. Connect the OPEN standard (included in the calibration kit) to the
other end of the measurement cable that is connected to the test
port 1 as shown in the following figure, and measure the open
calibration data at the test port 1. After measuring the open
calibration data, a checkmark is displayed to the left of the Port 1
Open menu.
Cal > Calibrate > 2-Port Cal > Reflection > Port1 Open
In the same way, measure the calibration data for the SHORT/LOAD
standards at the test port 1.
Connecting the OPEN/SHORT/LOAD standards

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals