EasyManuals Logo

Agilent Technologies E5071C User Manual

Agilent Technologies E5071C
2685 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #814 background imageLoading...
Page #814 background image
Programming
1041
Averaging Trigger Function
• Averaging Trigger
• Toggling Mode
Other topics about Making Measurement
Averaging Trigger
The averaging trigger function is used to execute the number of sweeps
specified as the averaging factor with a single trigger when the sweep
averaging function is used. When a trigger is given, the sweep count is
cleared, the sweep is executed the number of times specified by the
averaging count, and then the instrument goes into Hold status.
The setting of the averaging trigger function is valid for all channels. Note
that you can set the sweep averaging function for each channel.
Toggling Mode
To toggle ON/OFF the averaging trigger function, use the following
command:
:TRIG:AVER
• When the point trigger function is on, its setting has priority and
the setting of the averaging trigger is ignored. More specifically,
the required number of triggers is: number of measurement
points in a single sweep x averaging factor.
• When the averaging function is off for the active channel, the
averaging trigger function does not work for the channel. When
a trigger is given, the sweep is executed only once, and then
the instrument goes into Hold status.

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals