Programming
1073
36}:TRAN:TIME:KBES
Impulse width
:CALC{1-
36}:TRAN:TIME:IMP :WIDT
Rise time of step
signal
:CALC{1-
36}:TRAN:TIME:STEP
:RTIM
The above three items are dependent on each other. When the value of
one of them is changed, the values of the other two are automatically
changed to corresponding values.
Unlike manual operation, you cannot set the window shape by selecting
the window type (maximum/normal/minimum). However, you can set the
same shape as each window type by setting β as follows:
Maximum Normal Minimum
Value of β 13 6 0
Setting Display Range
To set the display range after time domain transformation, use the
following commands:
Description Command
Start value
:CALC{1-36}:TRAN:TIME:STAR
Stop value
:CALC{1-36}:TRAN:TIME:STOP
Center value
:CALC{1-36}:TRAN:TIME:CENT
Span value
:CALC{1-36}:TRAN:TIME:SPAN
Deleting unnecessary measurement data in the time domain
You can use the Gating Function to delete unnecessary time domain data.
On/Off
To turn on or off the gating function, use the following command:
:CALC{1-36}:FILT:TIME:STAT
Selecting Gate Type
To select the gate type, use the following command:
:CALC{1-36}:FILT:TIME

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals