E5071C
2788
:CALCulate:ATRaces:TIME:STEP:AMPLitude
Type
Command/Query
Syntax
:CALCulate:ATRaces:TIME:STEP:AMPLitude <numeric>
:CALCulate:ATRaces:TIME:STEP:AMPLitude?
Description
This command sets step amplitude value for the transform function.
Parameter
Description Stimulus amplitude level
Data Type Double precision floating point type (Double)
Range 1m to 5
Preset Value
200m
Unit volt
Resolution -
Query Response
{numeric}<newline><^END>
Examples
10 OUTPUT 717;":CALC:ATR:TIME:STEP:AMPL 100e-3"
20 OUTPUT 717;":CALC:ATR:TIME:STEP:AMPL?"
30 ENTER 717;A
Equivalent GUI
Setup > Stim. Ampl. (under Basic)