E5071C
2908
Performing Measurement
Calibration
Making Measurements
Other topics about Controlling Multiport Test Set
Calibration
Follow these steps to perform calibration with the multiport test set
connected:
1. Press Channel Next (or Channel Prev) key to set the channel that you
want to calibrate to the active channel.
2. Follow Displaying the properties to display the multiport test set
properties.
3. Perform the calibration. Check the corresponding multiport test set
test ports shown in the calibration properties as the port names of
the E5071C are displayed on the calibration menu, connect the
calibration standard to the corresponding test ports of the multiport
test set, and perform the calibration.
Making Measurements
Making measurements with multiport test set is the same as E5071C
standalone measurement. Perform operations by referring to Making
Measurements.
Trigger state and switching the setting of the Multiport test set
The following table shows how the setting in the multiport test set is
switched from when the trigger state is the stop state.
Trigger state Switching the setting of multiport test set
Stop The setting is not switched.
Trigger wait
The setting of the internal switch and the output of the control line
are switched according to the setting of the channel swept first.
The connections of the test ports and the output of the control line
are switched according to the setting of the channel swept first.
Measurement
Measurement is performed following the procedure below.
1. Execute a sweep for the first channel.
2. Set the connections of the test ports and the output of the
control line according to the settings of the channel swept