Controlling Multiport Test Set
2911
If you turn OFF the control of the multiport test set, it does not affect the
operation of the E5071C, even if it is connected. You can control test set 1
and test set 2 separately.
Selecting ports assigned to Port 1 to Port n
Selecting the connection ports
You can select the ports assigned to Port 1 to Port n for each channel. To
select the ports, use the following commands:
:SENS{1-36}:MULT{1-2}:PORT{1-20}
The connection between the assigned ports and Port 1 to Port n
inside the E5091A is not changed when one of the above
commands is executed but it is changed immediately before a
sweep for each channel.
Turning ON/OFF state display of connection ports (properties display)
You can display the state of the ports assigned to Port 1 to Port n
(multiport test set properties) in the lower right part in the window for
each channel. To turn ON/OFF the properties display, use the following
command:
:SENS:MULT{1-2}:DISP
Checking number of ports
To check the number of the multiport test set connected ports, use the
following command:
:SENS:MULT{1-2}:COUN?
Checking the available port name
To check the name of the multiport test set connected ports, use the
following command:
:SENS:MULT{1-2}:PORT{1-20}:CAT?
Checking number of input ports
To check the number of the multiport test set input ports, use the following
command:
:SENS:MULT{1-2}:INC?
Setting control line
You can set the HIGH/LOW state of each line of the control line for each
channel. To set the HIGH/LOW of each line, use the following command:
:SENS:MULT{1-2}:TSET9:OUTP
The HIGH/LOW state of each line of the E5091A is not changed
when the above command is executed but it is changed
immediately before a sweep for each channel.