Product Information
761
257 error command is not executed correctly. This error
occurs when you try to copy to an unsuitable file
name, for example.
-
256
File name
not found
The file name specified is not found and hence the
command is not executed correctly. This error
occurs when you try to read a file that does not
exist in a disk or a disk is not correctly inserted
into the drive to read or write a file, for example.
107
File
transfer
failed
This error occurs when writing data into or
reading data from a file (MMEM:TRAN command)
fails.
56
Fixture
Simulator
not
allowed
This error occurs when the fixture simulator is
executed (CALC:FSIM:STAT ON) under the
condition of the frequency offset is enabled
(SENS:OFFS ON).
The fixture simulator is allowed to use when the
frequency offset is not enabled.
24
Frequency
offset
enabled
This error occurs when the normal calibration
(SENS:CORR:COLL:OPEN etc.) is performed
under the condition of the frequency offset is
enabled (SENS:OFFS ON).
59
Frequency
offset not
allowed
This error occurs when frequency offset function
is not allowed in the current setting.
25
Frequency
offset not
enabled
This error occurs when the scalar-mixer
calibration (SENS:CORR:OFFS:COLL:OPEN etc.) is
performed under the condition of the frequency
offset is not enabled (SENS:OFFS OFF).
62
Frequency
out of
range
This error occurs when the setting of frequency
exceeds output frequency range under the
condition of the frequency offset is enabled
(SENS:OFFS ON).
G
No. Message Description
-105
GET not A group execution trigger (GET) has been
received in the program message (see 7.7,