4 OPERATION PROCEDURES CCW-R-2**
INSTRUCTION MANUAL
4-10
6. Scroll the screen by using the scroll key or the
scroll bar until the desired preset item is
displayed.
7. Press the [Preset Item] key to select the item.
8. Press the [OK] key .
X The [WH Zero Adjustment Start] screen of
the start-up assistant menu appears.
Fig.4-16 [C2 Preset Slct] Screen
(Start-up Assistant Function)
On the [WH Zero Adjustment] screen,
the weigh hoppers are displayed in
blue. However, they are displayed in
red right after power-on or when the
zero adjustment is not performed for 30
minutes or more. When the weigh
hoppers are displayed in red, it is not
possible to proceed to the next step until
the zero adjustment is performed.
9. Press the [Start] key .
Fig.4-17 [WH Zero Adjustment] Screen
(Start-up Assistant Function)
XZero adjustment operation for the weigh
hopper starts.
The progress bar is displayed during zero
adjustment.
When an error occurs during zero
adjustment, the weigh hoppers are
displayed in red. In this case, press the
[Start] key again and perform
zero adjustment.
X The [DF Zero Adjustment Start] screen of
the start-up assistant menu appears.
Fig.4-18 [In WH Zero Adjst.] Screen
(Start-up Assistant Function)