4 OPERATION PROCEDURES CCW-R-2**
INSTRUCTION MANUAL
4-18
<Zero adjustment for the dispersion table>
1. In the [Zero Adjustment] screen, press the [Slct
All DF] key .
XAll dispersion tables are displayed in blue.
2. Make sure that there is no product on the
dispersion table.
Fig.4-32 [Zero Adjustment] Screen
3. Press the [Start] key .
The [Start] key cannot be pressed when
the control unit power switch is OFF.
Fig.4-33 [Zero Adjustment] Screen
X The message “Please wait a moment.”
appears on the screen, and zero adjustment
starts.
Fig.4-34 [Zero Adjustment] Screen (In Adjustment)