4535 611 98931 iE33 Service Manual Page 106
CSIP Level 1 Installation: Site Inspection
.
1. Height with video monitor retracted and locked into position is 139.7 cm (55 in); height with monitor
extended is 166.5 cm (65.5 in); weight with peripherals is 177.3 kg (390 lb).
2. Atmospheric pressure must be between 1,060 hPa to 700 hPa (1.0 to 0.7 Atm).
3. Atmospheric pressure must be between 1,060 hPa to 500 hPa (1.0 to 0.5 Atm).
4. Peripherals will increase heat load. Discuss with the facilities engineer the possible requirement for
additional air circulation to handle the heat load from the system.
Table 5-1 System Specifications (Crated/Uncrated)
Parameter Metric Value U.S. Value
Height
1
166.5/139.7 cm 65.5/55 in
Width 57 cm 22.5 in
Depth 113.0 cm 44.5 in
Weight with Peripherals
1
161 kg 355 lb
Fully Crated Shipping Weight (VCR separate) 247 kg 545 lb
Ambient Operational Temperature
2
16°C to 40°C61°F to 104°F
Internal Operational Temperature
2
16°C to 55°C61°F to 131°F
Storage Temperature
3
-34°C to 65°C-29°F to 149°F
Storage Humidity 15% to 95% RH (non-condensing)
Heat Output
4
2,824 kJ/h at 785 W 2,677 BTU/h at 785 W