CSIP Level 1 Safety: Introduction
3Safety
Introduction
Please read this information before using your ultrasound system. It applies to the
ultrasound system, transducers, recording devices, and any optional equipment.This device is
intended for use by, or by the order of, and under the supervision of a licensed physician quali-
fied to direct the use of the device.
A WARNING describes precautions necessary to prevent injury or loss of life.
A CAUTION describes precautions necessary to protect the equipment.
Electrical
Safety
This equipment has been verified by a recognized third-party testing agency as a Class I device
with Type BF and Type CF isolated patient-applied parts and Type B accessories. For maximum
safety observe these warnings:
WARNINGS • Shock hazards may exist if this system, including all externally mounted recording and moni-
toring devices, is not properly grounded. Protection against electrical shock is provided by
grounding the chassis with a three-wire cable and plug. The system must be plugged into a
grounded outlet. The grounding wire must not be removed or defeated.
• Do not remove the protective covers on the system; hazardous voltages are present inside.
Cabinet panels must be in place while the system is in use. All internal adjustments and
replacements must be made by a qualified Philips Ultrasound field service engineer.
• Do not operate this system in the presence of flammable gases or anesthetics. Explosion can
result.