4535 611 98931 iE33 Service Manual Page 125
CSIP Level 1 Installation: Connectivity Information
Ultrasound
System
Configuration
➤ To configure the ultrasound system for a network connection
1. Power up the ultrasound system.
2. Press Setups (on the keyboard).
3. Click Print/Network.
4. Click the Global Config tab.
5. Click System.
6. Enter the following:
- System Name
- AE Title
- Port Number (a change in this field will cause the system to reboot)
- IP Address
- Subnet Mask
-Gateway
- Station Name
- System Location
- MAC Address (this is a derived address that is read from the iE33 network card)