4535 611 98931 iE33 Service Manual Page 256
CSIP Level 1 Troubleshooting: System Software Tools
3. Set the Acquisition\Channel Walk utility parameters as follows
-Channel
- Transmit
- Walk ā1ā
- Automatic
- Delay 0.5 sec
- Transducer Port 1
- SH Element to 1
4. While imaging a phantom, slowly scroll the tower(s) across image and verify the appearance
of the tower for each element is consistent with the towers for all other elements.
5. Select Receive.
6. While imaging a phantom, slowly scroll the tower(s) across image and verify the appearance
of the tower for each element is consistent with the towers for all other elements.
7. Click Stop.
8. Connect the transducer to the next port.
9. Set SH Port to this port.
10. Click > to begin scrolling.
11. Repeat step 4 through step 10 until all the ports have been tested.
12. Click Finish.
FEC Debug Port
The FEC debug port may be used to determine the configuration of the front end PCBs (includ-
ing part numbers), the version number of front end ASICs, and if there are faults in the Acquisi-
tion Subsystem. The detectable faults include power monitor faults and hardware configuration