4535 611 98931 iE33 Service Manual Page 477
CSIP Level 1 System Administration: System Admin
To export images to DVD
1. Press Review. The patient directory is displayed.
2. Select the images you want to export.
3. Click To DVD as JPG/AVI in the Review Exam display.
4. If the Format DVD dialog box is displayed, click OK, click Format in the Review Exam
display, and follow the prompts. Then click To DVD as JPG/AVI again.
5. In the Copy to DVD dialog box, select or create a directory in which to store the images
on the DVD.
6. Click OK to start copying. A message indicates when the images have been copied.
To export images in DICOM format to DVD
1. Press Review. The patient directory is displayed.
2. Click Select All. The images are highlighted with reverse video.
3. Click Send To. The Send To dialog box is displayed.
4. Select DVD.
5. Insert a DVD into the drive.
6. Click OK.
7. Remove the DVD when copying has completed.
Saving User
Settings
This procedure saves all Tissue Specific presets, system settings, analysis settings, annotation set-
tings, body markers, printer and network settings, audio/video settings, and voice profiles. The
settings must be saved and restored prior to and after re-imaging or upgrading.