4535 611 98931 iE33 Service Manual Page 234
CSIP Level 1 Preventive Maintenance: Operational Test and Paperwork
6. Type f and press Enter.
7. Type cd host and press Enter.
8. Type cd bin and press Enter.
9. Type dmmScript and press Enter. It takes several minutes to run the script.
10. Type cleanLsr.bat and press Enter.
11. Power down the system using the circuit breaker and wait several seconds.
12. Reboot the system.
Operational
Test and
Paperwork
After completing the PM procedure, operationally test the system by initiating a System Confi-
dence Test. Complete PM paperwork by filling out a Quality Assurance Label and a work order.