4535 611 98931 iE33 Service Manual Page 133
CSIP Level 1 Installation: Installation Completion
Print Image
Quality
➤ To adjust print image quality
1. Press Setups on the keyboard.
2. Click Print/Network.
3. Click the Printer/Capture tab. The Image Export dialog box is displayed.
4. Select the appropriate printer settings based upon customer expectations of printed image
quality.
NOTE The numbers in the boxes are Philips stipulated default settings. The customer may
choose other settings based upon personal preferences.
5. Use “trial-and-error” until the desired print settings are obtained.
6. Refer to the printer manufacturer’s conformance statement to determine the acceptable val-
ues for LUT (look-up table) items such as the numbers used for maximum or minimum den-
sity.
Installation
Completion
Final Inspection
1. Perform an inspection if one was not previously completed. Record all pertinent information
on the work orders.
2. Remove the protective sheet from the touch screen.
3. Verify the electrical power quality using the system and its peripherals as the final load. Also,
compare your pre-installation observations of ambient temperature, RFI, and ESD conditions
to present conditions. Record all measurements and observations on the Installation Work
Order.