4535 611 98931 iE33 Service Manual Page 198
CSIP Level 1 Performance Tests: Connectivity and Review
5. Acquire an image using Sincle Sweep 3D. Verify that a volume has been acquired.
6. Acquire an image using dynamic 3D and note the presence of a live volume. (Move the region
of interest box to the image data if necessary.)
Successful execution of the above steps indicates the appropriate licensed options are installed
Connectivity
and Review
To test connectivity and review functions
1. Press 2D and press Color.
2. Turn Color to adjust color gain until there is color noise in the color box.
3. Press Print. Verify that you hear a beep and the acquired image appears in the thumbnail on
the right side of the display.
4. Press Capture (while still in live imaging). Verify an hourglass appears and a delayed beep is
heard. Also verify that another thumbnail is displayed with a loop icon on the right side of the
display.
5. Press End Exam.
6. Press Review to display the Patient Directory.
7. Insert a new DVD into the DVD drive.
8. Click Format to format the disk.
9. Select the patient exam that was most recently ended and click Send To.
10. In the Send To dialog box, select DVD and click OK.
11. Verify that the green dot on the DVD queue icon comes on briefly.
NOTE Any other indication (orange or red dot) implies that there is a problem with the disk,
drive, or the IEP PCB. Troubleshoot as necessary.