CSIP Level 1 Performance Tests: 2D
4. Touch a clinical option and a Tissue Specific preset.
5. Touch Tra ns duce r.
6. Press Patient Data and enter a patient name.
7. Click Close.
Successful execution of the above steps implies:
Transducer ports are operational
Transducer, patient data, and the keyboard are operational
Touch screen is operational and that the calibration has been successful. No further testing
of the touch screen is necessary.
2D
To test 2D imaging
1. Touch 2D to enter 2D imaging (if needed).
2. Turn Gain to change the gain.
3. Adjust TGC slides. Verify changes in overall image brightness.
4. Turn Zoom. Verify the image increases from and decreases to the original size.
5. Turn Depth. Verify the changes in depth.
6. Press Zoom to bring up the HD Zoom box.
7. Use the trackball and Select control to move and size the HD zoom box.
8. Press Zoom to enlarge the boxed area.
9. Press Zoom to return to 2D imaging.
10. Adjust Res/Spd. Verify the icon updates on the touch screen.
11. Adjust Gray Map. Verify the map changes on the display and the map number changes on
the touch screen.