4535 611 98931 iE33 Service Manual Page 199
CSIP Level 1 Performance Tests: Annotations and Body Markers
12. Delete the patient that was copied to the DVD from the hard drive.
13. Select the DVD directory, highlight the exam that was copied, and transfer the exam to the
hard drive.
14. Select the hard drive directory and the exam that was copied from the DVD. Click on View
Selected. Verify that if in a 4:1 image presentation, the first loop appears as a JPEG loop
icon, while the print appears as it was taken.
15. Double-click on the JPEG loop icon to obtain a full-screen playback of that loop.
16. Touch Cine Play.
17. Click Close.
Successful execution of the above steps indicates
• End Exam and Review are working
• the patient database is functional
• DVD drive is operational
• Transfer of JPEG loops to the DVD and back means that the IEP board is functional after it
had been re-flashed during the SW installation
Annotations
and Body
Markers
➤ To test annotation and body marker functions
1. Press 2D and obtain a 2D image.
2. Press Annotate. Verify that the Annotation touch screen is displayed with a blinking green
curson on the main display.
3. Touch any annotation term on the touch screen. Verify that the annotation appears on the
main display at the cursor location.
4. Press Erase Text. Verify that annotation disappears from the display.