4535 611 98931 iE33 Service Manual Page 217
CSIP Level 1 Adjustments: Control Panel Articulation Adjustments
Control Panel
Articulation
Adjustments
To adjust the control panel articulation-cable tension
1. Adjust the control panel articulation trigger-release tension (Figure 7-7).
a. Raise the control panel articulation arm using the trigger release.
b. Remove the four screws securing the lower handle half to the upper and remove the han-
dle half.
c. Looking at the cable-adjustment nut from the non-trigger side of the cable, adjust the nut
clockwise to increase tension on the cable or counterclockwise to decrease tension.
d. Adjust tension to the customer’s preference.
e. Replace the lower handle half.
2. Adjust the control panel articulation tension at the base of the articulation arm (Figure 7-8).
a. Lower the control panel articulation arm using the trigger release.
b. Remove the top lid enclosure from over the base (Figure 10-9).
c. Using a wrench, adjust the nut clockwise to increase tension or counterclockwise to
decrease tension.
d. Adjust tension for operator preference.
e. Replace the top lid enclosure.
3. Using the articulation trigger release, check the articulation adjustment by moving the con-
trol panel up and down.