4535 611 98931 iE33 Service Manual Page 260
CSIP Level 1 Disassembly: Introduction
10 Disassembly
Introduction
WARNING Always turn off power, disconnect the main power cord from the wall outlet, and wait at least 30
seconds before removing or installing any PCB, module, or component.
CAUTION Always use correct ESD procedures. ESD damage is cumulative and may not be noticeable at
first. ESD symptoms may be first exhibited as a slight degradation of performance or image
quality.
NOTE The touch screen, video monitor, and all batteries inside this product contain mercury
and must be recycled or disposed of according to local, state, and federal laws. Do not
replace individual batteries in the field. Return all products containing mercury to Both-
ell for disposal.
Refer to Figure 10-3 and Figure 10-4 for links to the individual disassembly procedures.