4535 611 98931 iE33 Service Manual Page 231
CSIP Level 1 Preventive Maintenance: Monitor Articulation
Monitor
Articulation
The recommended interval for monitor articulation maintenance is every 12 months. Refer to
“To adjust video monitor elevation, tilt, and swivel” on page 204 if adjustment is necessary.
➤ To check monitor articulation
1. Check the monitor articulation mechanism for smooth operation.
2. Check that the monitor does not droop (tilt). Adjust the allen-head screws at the tilt point.
(Figure 7-2).
3. Check that the monitor holds a stable vertical position (buoyancy) during normal use. Adjust
the tension on the jack screw at the articulation joint (Figure 7-1).
4. Check the monitor swivel mechanism. Adjust the swivel nut at the articulation arm swivel
point (Figure 7-3).