4535 611 98931 iE33 Service Manual Page 227
CSIP Level 1 Preventive Maintenance: Control Panel Articulation
Control Panel
Articulation
The recommended interval for control panel articulation maintenance is every 12 months.
To check control panel articulation
1. Check the control panel articulation for smooth operation (up-down and left-right).
2. Check the control panel side-to-side release mechanism for proper operation (Figure 8-6):
a. Check control panel side-to-side movement. Adjust the tension if necessary (Figure 7-5).
b. Check the control panel swivel movement.
3. Check the control panel vertical-release lever for proper operation. Adjust the cable tension
on the release lever if necessary (Figure 7-7).
4. Check the control panel articulation-arm movement. Adjust the tension at the base of the
arm if necessary (Figure 7-8).
5. Check that the keyboard slide mechanism is fully functional. Adjust the slide catch if neces-
sary (Figure 7-6).
6. Verify that the keyboard light turns on when the keyboard is pulled out and turns off when
the keyboard is pushed in. Adjust the keyboard if necessary.